Baseline Calibration
A certified ceramic tile with known light reflectance across the visible spectrum establishes the baseline for spectrophotometer calibration. Positioning the white reference standard before the instrument’s optical port allows the system to calibrate its maximum reflectance scale. This process is required at the start of every shift in a textile testing lab.
Traceable Materials
The tile is usually made of sintered polytetrafluoroethylene or high-purity barium sulfate, which provides near-perfect diffuse reflectance. Each tile is calibrated by the manufacturer against a master standard maintained by a national standards institution. It comes with a stored digital file containing its exact reflectance percentages at each wavelength.
When the calibration run executes, the software compares the measured signal to this file to calculate the gain correction factors for each sensor channel.
Operational Care
Because any fingerprint, scratch, or dust speck will alter the light reflectance, the tile must be stored in a protective case and never touched with bare fingers. Cleaning must be performed using distilled water and lint-free optical wipes.
Measurement Limit
Calibrating the upper end of the reflectance scale does not correct for errors in the dark region of the spectrum. A separate step using a black trap or open-port measurement is required to complete the calibration process.