Vibrational Analysis
Molecular identification relies on the absorption of light at specific frequencies to detect functional groups within a fibre or finish. Utilization of infrared spectroscopy allows for the observation of the characteristic stretching and bending of chemical bonds in the mid-infrared range. This non-destructive test provides a graphical representation of the energy absorbed, known as a spectrum, which confirms the identity of unknown textiles.
Every polymer has a unique pattern that acts as a reliable identifier in forensic textile analysis and quality control.
Polymer Identification
Distinguishing between different types of nylon or polyester is simplified by comparing the sample peaks against a known library. Small changes in the chemical structure, such as the addition of a flame retardant or a moisture-wicking finish, appear as distinct shifts in the data. Hand-held units enable rapid screening of raw materials at the factory gate.
Quantitative Assessment
Measurement of peak intensity provides an estimate of the concentration of specific additives in a fabric blend. By applying the Beer-Lambert law, the lab can determine if a finish meets the specified levels required for performance claims. Surface treatments on technical textiles are frequently verified using this approach.
Spectral Interference
Proper sample preparation, such as drying the fabric or using an attenuated total reflectance crystal, mitigates the risk of moisture interference. Background noise from atmospheric CO2 is typically subtracted by the software to ensure the purity of the signal.